Siemens Mammomat 3000 Nova
We provide quality Refurbished Medical imaging Systems. We are refurbisher of excellent used Mammography systems. In refurbished Mammography systems, we provide Siemens Mammonat 3000 Nova/ 1000.
Technical Specifications
Generator
- High-voltage wave form: Multipulse
- Exposure voltage : 23kV to 35kV, adjustable in increments of 1kV
- mAs range : Mo X-ray tube: 2mAs to 560 mAs (mAs-mode)up to 600mAs (AEC-mode)
- Mo/W X-ray tube : 2mAs to 710 mAs (mAs-mode) up to 752 mAs (AEC-mode)
- Exposure times : From 10 ms to 4s , From 10 ms to 7s for magnification technique
Automatic Exposure Control
- Automatic exposure control (AEC): Microprocessor controlled, transparency compensation
- Detectors: Semi conductor detector
- Accuracy of optical density (2-6 cm): ± 0.15 OD from the mean optical density for predefined film screen combinations (separately for each grid table, magnification table and anode-filter combination at appropriate clinical kV).
- Density correction : ± 3 exposure points, adjustable in 1/8 step increments
- Film-screen combinations : 2; User selectable on the generator control panel (H/D)
X-Ray Stand
- Height adjustment: Motorized, 650 mm to 1350 mm floor to object table (in all angles)
- Swivel range: +135° / -180°
- Rotation of swivel arm: Motorized, isocentric
- Source to Image Distance (SID): 65 cm
- Film formats: 18cm x 24cm, 24cm x 30cm (15 mm thick Mammography cassettes according to DIN 6832/IEC 406). Automatic collimation to film format
- Grid: Reciprocating; Pb 4/27 (4:1 ratio, 27 lines/cm)
- Compression device: Opcomp, motorized compression initiated via footswitch or manual adjustment
- Compression force: Maximum motorized force 200N
- Magnification technique: Factor of 1.5 and 1.8
- Power supply: 110V, 208V, 230V, 240V, 277V, ± 10%, 1-phase; 208V, 230V, 277V, 400V ± 10%, 2-phase; 50 Hz or 60 Hz
- Fuse: 35A at 110V, 20A at 208V to 400V